While under contract to the United States Navy in the late 1940s,
Wallace H. Coulter developed a technology for counting and sizing particles using impedance measurements.
The technology was principally developed to count blood cells quickly by measuring the changes in
electrical conductance as cells suspended in a conductive fluid passed through a small orifice.
Presently, over 98% of automated cell counters incorporate this technology,
which is referred to as the Coulter Principle. In the past fifty years,
the technology has also been utilized to characterize thousands of different industrial particulate materials as well.
Beckman Coulter
® instrument systems which utilize this principle are called COULTER COUNTER
® instruments.
Drugs, pigments, fillers, toners, foods, abrasives, explosives, clay, minerals, construction materials,
coating materials, metals, filter materials, and many other sample types have all been analyzed using the
Coulter Principle. It can be used to measure any particulate material that can be suspended in an electrolyte solution.
Particles as small as 0.4 µm and as large as 1200 µm in diameter can routinely be measured.
Over 8,000 references to the uses of this technology have been documented.
In a COULTER COUNTER instrument, a tube with a small aperture on the wall is immersed into a beaker that contains
particles suspended in a low concentration electrolyte. Two electrodes, one inside the aperture tube and one outside
the aperture tube but inside the beaker, are placed and a current path is provided by the electrolyte when an electric
field is applied (Figure 1). The impedance between the electrodes is then measured. The aperture creates what is
called a "sensing zone". Particles in low concentration, suspended in the electrolyte, can be counted by passing
them through the aperture. As a particle passes through the aperture, a volume of electrolyte equivalent to the
immersed volume of the particle is displaced from the sensing zone. This causes a short-term change in the impedance
across the aperture. This change can be measured as a voltage pulse or a current pulse. The pulse height is proportional
to the volume of the sensed particle. If constant particle density is assumed, the pulse height is also proportional
to the particle mass. This technology thus is also called aperture technology.
Using count and pulse height analyzer circuits, the number of particles and volume of each particle passing
through the aperture can be measured. If the volume of liquid passing through the aperture can be precisely
controlled and measured, the concentration of the sample can also be determined. In modern COULTER COUNTER
instruments, such as Beckman Coulter’s Multisizer™4 particle counter and sizing instrument, pulses are
digitized and saved with several key parameters that describe each pulse such as pulse height, pulse width,
time stamp, pulse area, etc. These parameters enable the instrument to better discriminate between noise and
real pulses and between normal pulses and distorted pulses due to various reasons when particles pass through
the aperture. The saved pulses can be also used to monitor sample changes over the measurement time period if
pulses are arranged in time sequence. In practice, the particle volume is often represented in terms of equivalent
spherical diameter. The measured particle volume (or size) can be then used to obtain particle size distribution.
Figure 1. Schematic of a COULTER COUNTER.
A typical measurement using COULTER COUNTER instruments takes less than a minute, as counting and sizing rates of up
to 10,000 particles per second are possible. The accuracy of the size measurements can be better than 1%. Aperture
size typically ranges from 15 to 2000 µm. Each aperture can be used to measure particles within a
size range of 2 to 60% of its nominal diameter. Therefore, the overall particle size range of 0.4 to
1200 µm is feasible. However, the ability of the technology to analyze particles is limited to those
particles that can be suitably suspended in an electrolyte solution. The upper limit therefore may be
500 µm for sand but only 75 µm for tungsten carbide particles. Moreover, the lower size limit
is restricted by electronic noise generated mainly within the aperture itself. The selection of the most
suitable aperture size is dependent upon the particles to be measured. If the sample to be measured is composed
of particles largely within a 30:1 diameter size range, the most suitable aperture can be chosen. For example,
a 30 µm aperture can measure particles from about 0.6 to 18 µm in diameter. A 140 µm aperture
can measure particles from about 2.8 to 84 µm. If the particles to be measured cover a wider range than a
single aperture can measure, two or more apertures have to be used and the test results can be overlapped to provide
a complete particle size distribution.