| UNSPSC: |
56122003 |
| Autoclaving Cycles Conditions for Dispensing Cassettes: |
50 times |
| Autoclaving Pressure Conditions for Dispensing Cassettes: |
1 bar (15 psi) |
| Autoclaving Temperature Conditions for Dispensing Cassettes: |
121°C |
| Autoclaving Time Conditions for Dispensing Cassettes: |
20 min |
| Depth: |
32 cm (12.5 in) |
| Dispensing Accuracy: |
± 2% at 20 µL (typical) and ± 1% at 100 µL (typical) |
| Dispensing Precision: |
CV = 1.5% at 20 µL and CV = 1% at 100 µL |
| Dispensing Speed (384 Wells): |
16 seconds to dispense 20 µL into 384 wells |
| Dispensing Speed (96 Wells): |
5 seconds to dispense 20 µL into 96 wells |
| Dispensing Volume: |
5 to 100 µL (in 384 mode) |
| Fuses: |
Slow-blow type 2 x 600 mA for 115V and Slow-blow type 2 x 300 mA |
| Height: |
15 cm (6 in) |
| Number of Channels: |
8 |
| Operating Temperature Range: |
10°C to 40°C |
| Power Consumption: |
40 VA (measured at 230V/50 Hz) |
| Power Requirements: |
115-230 V ± 10%, 50-60 Hz |
| Storage Temperature Range: |
-25°C to 50°C |
| Weight: |
6.2 kg (13.6 lb) |
| Width: |
30 cm (12 in) |