For questions about products or ordering, contact us or call 1 800-526-3821 . Cart | beckmancoulter.com | sign in
Find: for this :


Home > CENTRIFUGATION > Products By Application > Genetic Analysis > Genetic Analysis-genomic DNA (debris cleanup) > Microfuge® 18 Microcentrifuge, Non-Refrigerated, with F241.5P Rotor, 60 Hz, 120 V

Microfuge® 18 Microcentrifuge, Non-Refrigerated, with F241.5P Rotor, 60 Hz, 120 V


Part Number: 367160



Includes F241.5P rotor.

Microfuge® 18 Microcentrifuge Features:
  • Maintenance-free rotor spins up to 14,000 rpm (18,000 x g)
  • Ideal for pelleting of DNA, RNA and Proteins; Cell Virus Isolation
  • Fits on crowded bench or in a cold room
  • Comes complete with a 24-place aerosol-tight, autoclavable fixed-angle rotor
  • Quiet operation <= 58 dB
  • Easy-to-read interface displays both speed (rpm) and force (rcf)
  • Runtime options - timed, hold or pulse
  • Easy-to-use knobs
Quantity:
 


Specifications

UNSPSC: 41103900
Height: 24.5 cm (9.7 in); with door open 39.5 cm (15.6 in)
Width: 24.0 cm (9.5 in)
Depth: 32.0 cm (12.6 in)
Weight: 13.0 kg (28.7 lb)
Approximate Acceleration Time: 18 seconds to max speed
Approximate Deceleration Time: 19 seconds from max speed
Electrical Requirements: 50 Hz, 230 VAC, 2 A; 50/60 Hz, 100 to 120 VAC, 4 A
Maximum Heat Dissipation into Room under Steady-State Conditions: 50 W (170 Btu/hr)
Noise Level 1 Meter in Front of Centrifuge: 58 dBA
Set Time: 0 to 30 min, continuous run (8), or pulse (short run)
Speed Range: 0 to 14,000 rpm
Ambient Temperature Range: 4 to 35°C ambient
Clearances Required: 7.6 cm (3.0 in)
Humidity Restrictions: < 80% (noncondensing)
Pollution Degree: 2
Ventilation Clearances Required: 7.6 cm (3.0 in) Sides and Rear