Optima™ L-80 XP Preparative Ultracentrifuge, 50 Hz, 220-240 VAC


Part Number: 392049



From the world leader in centrifugation comes the most intelligent ultracentrifuge system ever.

Whether your research involves proteomics, cytomics or genomics, our new Optima™ L-XP Series ultracentrifuge is designed to play a key role in the discovery process. Thanks to a sophisticated on-board computer, enhanced firmware, powerful Optima eXPert™ software and an easy-to-use touch screen, Optima L-XP delivers superior functionality, optimized user efficiencies, far greater capabilities and faster turnarounds than any other ultracentrifuge. In today's rapidly-changing discovery world, the Optima L-XP is engineered to put you on the inside track to tomorrow's exciting new breakthroughs.

The Optima L-XP Series are a new class of enhanced preparative ultracentrifuge instruments, rotors, tubes, accessories, and software, designed, manufactured and tested as "systems" from the inside and out to assure the highest quality separation in the shortest possible time, safely and reliably. These systems feature an air-cooled, vacuum-encased, frequency-controlled, imbalance-tolerant drive for fast and easy visual balancing of tubes, direct rotor inertia speed check, a 10-year drive warranty, and thermoelectric temperature control that uses no chlorofluorocarbons.

Efficiency, versatility, and environmental safety are the hallmarks of the Optima L-80 XP Ultracentrifuge from BeckmanCoulter. With Beckman Coulter’s innovative Optima centrifugation technology, the L-80 XP is capable of generating 602,000 x g at speeds of up to 80,000 rpm.The L-80 XP, along with a broad range of rotors, tubes, and supplies shortens run times for a wide range of applications. The L-80 XP offers the reliability of a vacuum-encased induction drive,Optima eXPert™ software, and environmentally-friendly cooling systems that eliminate the need for CFCs. And the L-80 XP’s special design eliminates the need for chlorofluorocarbons or other ozone-depleting coolants for drive and chamber cooling. CE/CSA/NRTL.

Quantity:
 


Specifications

Maximum Heat Dissipation into Room under Steady-State Conditions: 1.0 kW (3,400 Btu/hr)
Noise Level 1 Meter in Front of Centrifuge: < 53 dBA
Set Speed: Actual rotor speed ± 10 rpm of set speed after equilibration (above 1,000 rpm)
Set Temperature: 0 to 40°C in 1°C increments
Set Time: To 999 hr 59 min; Hold for continuous runs
Speed Range: 1,000 to 80,000 rpm
Temperature Control: ± 0.3°C of set temperature (after equilibration)
Ambient Temperature Range: 10 to 40°C ambient
Clearances Required: 5.1 cm (2.0 in) sides; 15.2 cm (6.0 in) rear
Humidity Restrictions: < 95% (noncondensing)
Pollution Degree: 2
Time -- Actual Display: Indicates time remaining in timed runs, time elapsed in Hold runs, or estimated time remaining in ?2t runs
Vacuum: Below 5 microns (0.7 Pa)
Finish: Control panel: coated polycarbonate. Top surface: urethane paint. Other surfaces: general purpose paint.
Installation Category: Category II
Display: Shows the accumulated ?2t to two significant digits (in exponential notation)
Top Surface: Urethane paint
Keypad Material: Coated polycarbonate
Manual Operation: Indicates time remaining in timed runs, time elapsed in Hold runs, or estimated time remaining in ?2t runs
Set ω²t: Up to 9.99 × 10^14 radians squared per second
Temperature Display: Actual rotor temperature in 0.1°C increments
UNSPSC: 41103907
Height: 125.7 cm (49.5 in)
Width: 94.0 cm (37 in)
Depth: 67.3 cm (26.5 in)
Weight: 465 kg (1,025 lb)
Approximate Acceleration Time: 10 acceleration profiles--9 slow rates or maximum acceleration from 0 to set speed
Approximate Deceleration Time: 11 deceleration profiles--9 slow rates, coasting to a stop without brake, or full dynamic braking from set speed
Electrical Requirements: 50 Hz, 220-240 VAC
Imbalance Tolerance: ±5 ml or 10%, whichever is greater