For questions about products or ordering, contact us or call 1 800-526-3821 . Cart | beckmancoulter.com | sign in


Home > PARTICLE CHARACTERIZATION > All Products > SA3100 with Dewar Lift -- Adsorption and Desorption

SA3100 with Dewar Lift -- Adsorption and Desorption


Part Number: 6605695



A complete solution for characterizing the surface of solid materials. The SA 3100 utilizes the well-established gas adsorption technique to measure the surface area and porosity of solids. Everything you need is incorporated into one powerful and easy-to-use package.

The SA 3100 with automated dewar lift is a complete systems with integrated outgassing stations and vacuum pump and is highly accurate. It features continuous saturation pressure measurement, has a minimum manifold volume to enhance sensitivity, and performs a multipoint helium free space measurement, to improve repeatability and accuracy.

Rapid Analysis is achieved through the use of optimized sample holders, greater sensitivity allowing a smaller sample volume and a novel learn function. It provides flexible output.

Data may be printed in full detail or in summary form:
  • Single and Multipoint BET data and Langmuir surface area data
  • Complete adsorption and desorption isotherm capability
  • B.J.H. pore size analysis
  • MP and T-Plot methods
SA 3100 Surface Area and Pore Size Analyzer

    Provide single and multipoint surface area analysis, multigas capability for surface area and full adsorption capability.

SA-VIEW™ Software

    This utility program has a Windows interface for an external PC and provides data reprinting, archiving, graphic overlays and other features.

Quantity:
 


Specifications

UNSPSC: 41113903
Depth: 53.3 cm (21")
Height: 71.1 cm (28 in)
Weight: 80 lb (including vacuum pump)
Width: 50.8 cm (20 in)
Power Requirements: 90 to 250 VAC, 50/60 Hz, 150W
Gas Purity Requirement - Adsorbates: 99.9% or better
Gas Purity Requirement - Helium: 99.995%
Manifold Temperature: 45 °C (± 0.1°C)
Mesopore Size Distribution Range: 3 nm to 200 nm
Micropore Analysis: Micropore Volume and Area by t-Plot Analysis
Minimum Measureable Relative Pressure: 0.00006
Minimum Sample Pressure Resolution: 0.046 mmHg
Operating Altitude: Up to 10,000 above MSL
Operational Power Consumption: 500W nominal
Operational Temperature Range: 10 °C to 35 °C
Operational/Storage Relative Humidity: 10% - 85% RH non-condensing
Outgas Furnace Settability: 1 °C
Outgas Furnace Stability: 5 °C
Outgas Furnace Temperature Accuracy: ± 5 °C
Outgas Furnace Temperature Range: 30°C to 350 °C
Outgas Ports: 3
Overall Pressure Range: Vacuum to 1000 mmHg
Specific Surface Area Range: From > 0.01 m²/g, dependent upon sample density
Storage Temperature Range: 0°C to 50 °C
Transducer Linearity: < 0.1% BFSL
Transducer Repeatability: 0.0125% BFSL
Typical Reproducibility: < 2% CV
Ultimate Vacuum: 10¯³ mmHg
Vacuum Pump Power Consumption: 130 W
Vacuum Pump Volume Displacement: 1.5 m³/hr