For questions about products or ordering, contact us or call 1 800-526-3821 . Cart | beckmancoulter.com | sign in


Home > SPECTROPHOTOMETRY > nanoVette Microliter Measurement > DU® 800 UV/Vis nanoVette Spectrophotometer, Life Science Package, without IBM PC System

DU® 800 UV/Vis nanoVette Spectrophotometer, Life Science Package, without IBM PC System


Part Number: A49423



DU® 800 Life Science Package - without IBM PC System

The DU 800 UV/Visible Spectrophotometer is a scanning instrument with a bandwith of less than or equal to 1.8 nm and a wavelength range from 190 to 1100 nm.

The DU 800 is a customer-installed spectrophotometer and System Installation should be ordered separately (except when a promotion includes service installation).

Instrument includes:
  • DU 800 Windows XP-Based System Control
  • RediScanâ„¢
  • RediReadâ„¢
  • Wavelength Scan
  • Fixed Wavelength (up to 12 wavelengths)
  • Kinetics/Time (12 reactions)
  • Single Component Analysis
  • Nucleic Acid Analysis (Ratios)
  • Performance Validation
  • Full Support of 21 CFR Part 11 (Electronic Records; Electronic Signature)
  • Start-Up Diagnostics.
  • nanoVette with 0.2 mm pathlength lid
Package includes: DU 800 UV/Vis Spectrophotometer, Protein Assay Analysis Software Option, Nucleic Acid Analysis II Software Option, Single Cell Holder, Single Cell Holder for 50 µL Microcell, 50 µL Microcell (Qty. 2).

Additional software applications are available and may be added to the system. (See Software Options).

When the PC is not ordered with the spectrophotometer, it is the responsibility of the customer to furnish a compatible PC with monitor and to supply and install Microsoft* Windows* XP as well as the DU 800 System and Applicaitons Software, which is included with the spectrophotometer.

Quantity:
 


Specifications

UNSPSC: 41115406
Depth: 59 cm (23 in)
Height: 53 cm (21 in)
Width: 69 cm (27 in)
Weight: 33 kg (73 lb)
Power: 200 watts typical
Ambient Temperature Range: +15 to 40°C (59 to 104°F)
Data Collection Rate: 20 Samplings per Second
Frequency: 50/60 Hz
Humidity Restrictions: < 85% maximum relative humidity, not to exceed 32.5°CWBT
Line Voltage: 100-240V ± 10%
Photometric Accuracy: ± 0.005A (at 1 A with NIST 930D solid filter at 546 nm)
Photometric Readout: -0.300 to 3.000A or 0.0 to 200%T
Response Time: 0.05 Seconds
RMS Baseline Flatness: From 200 to 900 nm, at 0 A
RMS Noise: < 0.0002A RMS (at 0A, average of 10 standard deviations of 10 readings at 0.05 sec intervals, at 500 nm)
Scan Speeds: 120, 240, 600, 1200, 2400 nm/min
Spectral Bandwidth: = 1.8 nm (from 200 to 680 nm)
Stability: < 0.003A/hr (0 A, constant ambient conditions, measured for one hour at 340 nm after one hour warm-up)
Stray Light: < 0.05% (measured using NaI at 220 nm, per ASTM E387-84)
Wavelength Accuracy: ± 0.2 nm at 656.1 nm, ± 0.5 nm at Full Range
Wavelength Range: 190 to 1,100 nm
Wavelength Repeatability: ± 0.1 nm at 656.1 nm, ± 0.2 nm at Full Range
Wavelength Setability: 0.1 nm Increments