Fast Response with Low-Ionic Strength Samples Beckman Coulter
STAR electrodes have been specially designed using a unique glass formula to
provide rapid response for samples with low ionic strength and when sample
temperatures differ significantly.
The following performance specifications will be met:
| Response Time: |
25ºC to 75ºC in pH 4 buffer; stable to ± 0.02 pH
in less than 30 seconds. With pH 3 to pH 7 buffer at constant
temperature; stable to ± 0.01 pH in less than 30 seconds |
| Accuracy: |
Relative to NIST certified buffers; ± 0.01 pH |
| Slope: |
97 to 100% |
| Isothermal Point: |
7.0 pH |
| Stability: |
± 0.002 pH over 24 hours |
|
|
STAR Combination Electrodes for Low-Ionic Strength Samples
|
|
|
|
Standard-length, refillable electrodes for measurements of samples with low ionic strength. STAR electrodes also provide a quick response to changing temperature. Available in either rugged epoxy body to protect the bulb or easy-to-clean glass body for use with all solvents.
|
| Part No. |
Description |
| 511070
|
Star™ Refillable Combination pH Electrode, Epoxy
|
| 511071
|
Star™ Refillable Combination pH Electrode, Glass
|
|
|
|
|
Silver Billet Electrode for Potentiometric Titrations and Acid Fluorides
|
|
|
|
Indicator electrode for potentiometric titrations and for the measurements of halides, silver and sulfides.
Electrode is provided with an attached cable (1 m / 3 ft.).
|
| Part No. |
Description |
| 511121
|
Silver Billet Reference Electrode with Attached Cable and Pin Connector
|
|
|
|
|
Electrode for Oxygen Reduction Potential (ORP) Measurements
|
|
|
|
Refillable Platinum combination electrode with a ceramic junction to measure Oxidation Reduction Potentials (ORP).
Electrode is provided with an attached cable (1 m / 3 ft.).
|
| Part No. |
Description |
| 511120
|
ORP Electrode with BNC Connector
|
|
|
|
|