|
|
| |
|
|
P/ACE MDQ Capillary Electrophoresis System for Quality Control
|
A CE-based analytical system configured optimally for the quality
assurance laboratory. With a focus on repetitive robust analyses, the
system includes a P/ACE MDQ configured with a filter-based UV/Vis (200,
214, 254 and 280 nm included) detector, UV source optics, ambient sample
storage module, large volume buffer reservoirs, and 32 Karat™ Software
configured on an IBM personal computer. Installation Qualification,
Operation Qualification 1 (OQ1) and documentation to aid in software
validation is also included.
|
|
Product Features
|
|
|
-
Capillary Thermoregulation with Recirculating Liquid Coolant
Recirculating liquid coolants dissipate "excess" heat much more
effectively than air-cooled systems. Efficient heat dissipation is
critical in capillary electrophoresis to allow for the usage of higher
conductivity buffers and larger I.D. capillaries. Your benefit is more
flexibility in sample load and the ability to create better separation
efficiencies with the use of higher ionic strength buffers. This allows
methods to be easily transferred from a methods development laboratory
to a routine use environment.
-
Multi-format Sample Introduction
Automation is certainly a means by which many laboratory bottle necks
may be reduced. The use of 96 well plates as a sampling vessel not only
increases system capacity but allows compatibility with automated sample
preparation devices which utilize the 96 well plate format.
Additionally, 2 ml vials (which are autosampler industry standards) may
be used; or sampling can be achieved directly from PCR vials and
microfuge tubes.
-
High Sensitivity UV Detector
Detector sensitivity in capillary electrophoresis is critical since
small mass loads are introduced and small detection pathlengths are used
(governed by the capillary dimensions i.e. 25 - 200 um). Fixed
wavelength detection in a filter wheel assembly maximizes detector
sensitivity yet still allows wavelength changes during a run. In a Q.C.
or regulated environment it is beneficial to "lock-down" a method once
it is developed in order to retain the integrity of the data. The P/ACE™
MDQ high sensitivity UV detector may be configured with a "single"
wavelength from a filter traceable to NIST making both the instrument
certification and method validation much simpler. Additionally the
system employs complete fiber-optic technology bypassing the necessity
of an optical bench. This lowers the noise on the detection system (by
removing reflective surfaces where energy loss can occur) and simplifies
the system making it more rugged for routine use. Diode array is an
essential detection tool for methods development work, but will reduce
the lifetime of a "coated" capillary. Therefore, once a method is
developed and is in routine use it is beneficial to switch to a UV
detector where the capillaries will last longer.
-
Multiple Separation Modes
Voltage gradient programming is beneficial for gel-sieving techniques
improving separations over a wide range of fragment sizes. This is
especially critical for the separation of nucleic acids. The application
of simultaneous voltage and pressure is beneficial for the high
efficiency mobilization of proteins during iso-electric focusing as well
as detecting material not migrated off the capillary. Also, several CEC
methods have specified the use of simultaneous pressure applied to both
inlet and outlet reservoir. This system is compatible with those methods.
-
Automatic Standards Co-Injection
Sample injections can be automatically spiked with reference standards
for positioning or for positive identification. This is useful to help
minimize errors resulting from sample matrix effects. Additionally, with
high salt samples it may be necessary to do multiple injections within
the same run in order to employ an electrokinetic salt elution step.
Practical examples of this include the electrokinetic injection of PCR
samples directly after amplification. The interfering salts and metals
would need to be be removed through a sample handling step if one were
not able to perform an electrokinetic elution from ddH20 prior to sample
introduction. Additionally, once a sample table has begun the data that
is generated may indicate the necessity to run new sample combinations.
On the fly sample table editing may be accomplished with the 32 Karat
Software allowing the addition of new samples to a running sequence.
-
Large Volume Buffer Reservoirs
When a method is validated for routine use, one's need changes away from
buffer type scouting to buffer capacity as a large number of samples are
processed using a single condition. With P/ACE™ MDQ, the buffer array
and reservoirs are interchangeable. Each buffer reservoir contains two
30 ml chambers for the inlet and outlet buffer and ten 2 ml vial
positions for conditioning fluids. The large volume reservoir allows
hundreds of runs to be facilitated from a single buffer tray alleviating
the need to frequently change buffers.
-
Electrokinetic, Pressure and Vacuum Sample Introduction
In moving validated methods from a development lab to a routine use
environment, the issue of compatibility of methods transfer is always
present. The P/ACE™ MDQ is designed to operate with all commonly used
sample introduction modes and has the capability to dial in the
parameters (i.e. injection pressure or vacuum), allowing greater ease in
transferring methods between instrument brands. Being able to inject
from the short side of the capillary, offers the capability of high
speed separations.
-
Fluid Delivery with Variable Pressure and Vacuum
Conditioning of the capillaries is accomplished by moving specific
volumes of electrolytes, regenerants and cleaning solutions through the
capillary. The P/ACE™ MDQ is designed to operate with all commonly used
rinsing protocols and the capability to regulate these parameters (i.e.
magnitude of applied pressure or vacuum), allows greater ease in
transferring methods.
-
GLP/CGMP Compliance Features
Many laboratories are mandated by law to follow current good
manufacturing practices (CGMP) and good laboratory practices (GLP) The
32 Karat Software has many features which are geared towards the
maintenance of data integrity. Sample data and methods parameters cannot
be overwritten or overridden, ensuring the data that was generated was
as described. To track instrument operation, all events including errors
are logged onto the system. To provide security of operation users may
be coded to different security levels with selected access to: system
configuration, methods programming, data reprocessing, specific
instrument operation and data comparison. All of these features can be
used to "lock down" the system for regulated use.
-
Specialized CE Calculations
Capillary electrophoresis requires specialized data handling. Mobility
calculations establish the identity of a compound independent from
changes in electroosmotic flow, voltage, temperature and column
dimensions. This can be very useful in compound identification and
method validation especially when comparing data between different
laboratories where analytical conditions may vary slightly. While
normalizing the peak area to analyte velocity is important if
quantitation is required. Intelligent system suitability, allows
unattended operation for routine use environments.
|
|

|
|
|
|
|
|
|
|