Specification |
Optima MAX-XP |
| Maximum speed (RPM) |
150,000 (with MLA-150 rotor) |
| Maximum force (RCF x g) |
1,019,000 (with MLA-130 rotor) |
| Speed control |
±50 rpm of set speed |
| Sample imbalance tolerance |
Up to 10% of volume in opposite tubes |
| Drive type/cooling air-cooled drive type |
Variable frequency induction drive, rotor overspeed ID |
| Refrigeration system |
Solid state, thermoelectric temperature
control system with forced air, no coolant,
no CFCs/ODCs |
| Set temp range |
0°C to 40°C (1°C increments) |
| Temp control |
±2°C of set with peak to peak temperature
fluctuation of 0.1°C |
| Ambient operating range |
15°C to 35°C |
| User-settable programs |
All user programs have up to five steps each |
| Acceleration/Deceleration |
10 accel profiles
11 decel profiles |
| Moisture-purging vacuum system |
Yes |
Power usage/heat output
steady state BTU/hr (kW) |
2,400 (0.7) |
| Sound pressure1m away |
<47 dBA |
| Sound power |
1.33 microwatts (A weighted) |
| Zwicker loudness |
6 sones |
| Dimensions WxDxH inches (mm) |
29 x 23 x 16.1 (737 x 584 x
409) |
| Weight lb (kg) |
230 (105) |
† BioSafe is a term intended to describe the enhanced Biosafety features of our products.