Beckman Coulter, Inc.
 

Products

   
   
 
 

Optima™ MAX-XP Benchtop Ultracentrifuge--
Specifications and Ordering Information

Printer Friendly


Related Product Information

Optima™ MAX-XP Benchtop Ultracentrifuge

Specifications

Specification
Optima™ MAX-XP
Maximum speed (RPM)

150,000 (with MLA-150 rotor)

Maximum force (RCF x g)

1,019,000 (with MLA-130 rotor)

Speed control

±50 rpm of set speed

Sample imbalance tolerance

Up to 10% of volume in opposite tubes

Drive type/cooling air-cooled drive type

Variable frequency induction drive, rotor overspeed ID

Refrigeration system

Solid state, thermoelectric temperature
control system with forced air, no coolant,
no CFCs/ODCs

Set temp range

0°C to 40°C (1°C increments)

Temp control

±2°C of set with peak to peak temperature
fluctuation of 0.1°C

Ambient operating range

15°C to 35°C

User-settable programs

All user programs have up to five steps each

Acceleration/Deceleration

10 accel profiles
11 decel profiles

Moisture-purging vacuum system

Yes

Power usage/heat output
steady state BTU/hr (kW)

2,400 (0.7)

Sound pressure1m away

<47 dBA

Sound power
1.33 microwatts (A weighted)
Zwicker loudness

6 sones

Dimensions WxDxH inches (mm)
29 x 23 x 16.1 (737 x 584 x 409)
Weight lb (kg)
230 (105)

Ordering Information

Part No.

Description

393315 Optima MAX-XP Benchtop Ultracentrifuge Electrical requirements: 220/240 VAC~, 6A, 50 Hz;120VAC~,12A,60Hz;100 VAC~, 12 A, 50/60 Hz
A47882 Optima MAX-XP Biosafe† Ultracentrifuge
includes Optima MAX-XP HEPA filter kit
393395 Remote monitoring and control option

† BioSafe is a term intended to describe the enhanced Biosafety features of our products.