| 
            
             | 
            Fully automated, random access chemistry system with STAT capability | 
        
        
            | 
            
             | 
            
             Spectrophotometry and potentiometry 
             | 
        
        
            | 
            
             | 
            
             Endpoint, rate, fixed point and indirect ISE 
             | 
        
        
            | 
            
             | 
            
             Colorimetry, turbidimetry, latex agglutination, homogeneous EIA, indirect ISE 
             | 
        
        
            | 
            
             | 
            
             60 photometric tests + 3 ISE, 120 pre-programmed onboard tests 
             | 
        
        
            | 
            
             | 
            
             400 photometric tests/hour; maximum of 800 with ISE 
             | 
        
        
            | 
            
             | 
            
             200 samples/hr. Maximum tests/hr: 600 if ISE only 
             | 
        
        
            | 
            
             | 
            
             Serum, plasma, urine, other 
             | 
        
        
            | 
            
             | 
            
             Racks with 10 samples each (barcodes on primary tubes and on racks); capacity of 80 samples; continuous loading 
             | 
        
        
            | 
            
             | 
            
             Primary and secondary tubes; diameter from 11.5 to 16 mm; height from 55 to 102 mm, nestled micro sample cups 
             | 
        
        
            | 
            
             | 
            
             Up to 22 positions for STAT samples via cooled turntable. Easy-to-operate interrupts between tests. Auto-repeat run capability. 
             | 
        
        
            | 
            
             | 
            
             1–25 μl in 0.1 μl steps (1-25 μl for repeats) 
             | 
        
        
            | 
            
             | 
            
             76 positions for (R1+R2, detergent position). Holds 15, 30 and 60ml bottles. 
             | 
        
        
            | 
            
             | 
            
             Refrigerated (4–12°C) 
             | 
        
        
            | 
            
             | 
            
             R1: 10–250 μl; R2: 10–250 μl (in 1 μl increment) 
             | 
        
        
            | 
            
             | 
            
             90–350 μl 
             | 
        
        
            | 
            
             | 
            
             Permanent glass cuvettes 
             | 
        
        
            | 
            
             | 
            
             Up to 8 minutes, 37.5 seconds 
             | 
        
        
            | 
            
             | 
            
             37°C, dry bath 
             | 
        
        
            | 
            
             | 
            
             Direct assay through the reaction cuvette (0–3.0 OD) mono and bichromatic measurements possible 
             | 
        
        
            | 
            
             | 
            
             13 different wavelengths between 340–800 nm 
             | 
        
        
            | 
            
             | 
            
             Auto calibration, cooled calibrator positions; master calibration established by two-dimensional barcode 
             | 
        
        
            | 
            
             | 
            
             Auto QC, cooled QC positions 
             | 
        
        
            | 
            
             | 
            
             1. Auto rerun and manual rerun available 
            2. Automatic sample increase, decrease or normal repeat 
            3. Reflex testing 
             | 
        
        
            | 
            
             | 
            
             Dilution ratio: Fixed to 3, 5, 10, 15, 20, 25, 50, 76 or 100 
             | 
        
        
            | 
            
             | 
            
             Individual and profile test requisition via online, mouse, function keys and touch-screen 
             | 
        
        
            | 
            
             | 
            
             Lipaemia, Haemolysis and Icterus analysis. Sample clot detection and probe crash protection. 
             | 
        
        
            | 
            
             | 
            
             RS232C. Full uni- and bidirectional communication possible. 
             | 
        
        
            | 
            
             | 
            
             Windows 7® 
             | 
        
        
            | 
            
             | 
            
             Sample number 100,000 samples. Reaction Data 200,000 tests. 
             | 
        
        
            | 
            
             | 
            
             ANL: 1450 mm x 1205 mm x 770 mm 
             | 
        
        
            | 
            
             | 
            
             100–240v; 60 Hz/ < 3.5 kVA 
             |